Clinical use: what matters at the bedside (physics-linked)
- Goal is to deliver sufficient myocardial current to terminate VF/pulseless VT; success relates more to delivered current through the heart than the selected energy on the machine.
- For a given selected energy, higher transthoracic impedance (TTI) reduces peak current and total charge delivered.
- Minimise TTI: good pad contact, correct position, firm pressure (paddles), adequate gel, avoid air gaps, shave hair if needed, use large pads, avoid placing over devices/metal.
- TTI falls with repeated shocks (skin heating, improved contact) and may be lower at end-expiration.
- Pad position aims to maximise current path through myocardium (anterolateral or anteroposterior).
- AP placement may reduce TTI and improve current through atria/ventricles in some patients; evidence mixed but commonly used for cardioversion.
- Biphasic defibrillators compensate for impedance (adjust voltage/pulse width) and achieve similar efficacy at lower selected energies than monophasic.
- Typical adult: biphasic 150–200 J (manufacturer-specific) vs monophasic 360 J.
- Synchronised cardioversion: shock timed to R wave to avoid delivery during vulnerable period (T wave) which can precipitate VF.
- If sync markers unreliable (e.g., very fast AF), increase gain/lead selection; if unstable, may need unsynchronised shock.
- Safety: oxygen away from chest, no one touching patient/bed, remove transdermal GTN patches, avoid wet surfaces; consider implanted devices and pacemakers.
- Place pads at least ~8 cm from pacemaker/ICD generator; use lowest effective energy; check device function afterwards.
Core definitions and quantities (exam essentials)
- Defibrillation: delivery of a brief high-energy electrical pulse to depolarise a critical mass of myocardium, extinguishing re-entrant wavefronts and allowing organised rhythm to resume.
- Energy (J): work done. In a capacitor discharge: E = 1/2 C V^2.
- For a fixed capacitor C, increasing selected energy requires higher charging voltage V (since E ∝ V^2).
- Charge (Q, coulombs): Q = C V. Delivered charge depends on waveform, pulse width and impedance.
- Current (I, amps): I = V/R (Ohm’s law). In defibrillation, R is mainly transthoracic impedance (TTI).
- Power (W): rate of energy transfer. P = V I = I^2 R = V^2/R (instantaneous).
- Current density (A/m^2): current per unit electrode area; relates to skin burns and local tissue injury.
- Larger pads reduce current density at skin and reduce TTI, improving myocardial current for a given energy.
Defibrillator circuit model (capacitor discharge through patient)
- Simplified model: capacitor (C) charged to voltage (V0) discharges through patient impedance (R) via switching circuitry → exponentially decaying current/voltage (monophasic damped sinusoid or truncated exponential depending on design).
- Time constant τ = R C (for simple RC exponential). Higher R increases τ (slower decay) but reduces peak current (I0 = V0/R).
- Modern biphasic defibrillators: controlled truncated exponential or rectilinear biphasic waveforms; microprocessor adjusts pulse width/voltage based on measured impedance.
- Energy actually delivered to patient is less than stored energy due to internal resistance and switching losses; displayed “selected energy” refers to intended delivered energy (device-dependent).
Waveforms: monophasic vs biphasic (physics and implications)
- Monophasic: current flows in one direction only.
- Types historically: damped sinusoidal (older) and monophasic truncated exponential (MTE).
- Higher energies required (commonly 360 J) and more myocardial injury/skin burns compared with biphasic for similar efficacy.
- Biphasic: current reverses direction part-way through shock (two phases).
- Mechanism: first phase depolarises; second phase helps remove residual membrane charge and reduces post-shock arrhythmogenicity, improving defibrillation threshold.
- Lower energy for similar success; less myocardial damage; better performance across a range of impedances due to impedance compensation.
- Pulse duration matters: too short may not capture enough myocardium; too long increases energy dissipation and injury. Devices choose pulse width to optimise delivered current/charge for measured impedance.
Transthoracic impedance (TTI): determinants and consequences
- TTI is the effective resistance to current flow between electrodes across chest wall and thorax; typical adult range ~50–100 Ω (can be wider).
- Higher TTI → lower peak current and reduced probability of successful defibrillation at a given selected energy.
- Factors increasing TTI: small pads, poor contact/air gaps, dry gel, hair, obesity, emphysema/hyperinflation, incorrect placement, low paddle pressure, transdermal patches, chest wall oedema.
- Factors reducing TTI: larger pads, good conductive gel, firm pressure, correct placement, shaving hair, repeated shocks, end-expiration.
Energy selection and what the number means
- Selected energy (J) is not the same as myocardial energy; what matters is myocardial current/charge delivered during the shock.
- For a given selected energy, high impedance reduces current; biphasic devices adjust waveform to mitigate this, but very high impedance can still reduce effectiveness.
- Defibrillation threshold (DFT): minimum energy (or current) required to achieve defibrillation with a given probability; varies with patient factors, electrode position, impedance and waveform.
- Biphasic waveforms lower DFT compared with monophasic.
Skin burns and tissue injury (physics)
- Burn risk relates to current density and contact quality: J = I/A (conceptually). Small contact area or air gaps concentrate current and heat.
- Paddles with firm pressure reduce impedance and improve contact area; adhesive pads provide consistent contact but must be applied correctly.
Synchronisation and the vulnerable period
- Unsynchronised shocks can fall on the T wave (relative refractory period) causing VF (R-on-T phenomenon).
- Synchronised cardioversion uses ECG sensing to time discharge with the R wave; requires adequate signal and correct lead selection.
Electrical safety and hazards (FRCA physics angle)
- Defibrillators are high-energy sources; risk to staff is mainly macroshock if in contact with patient/bed during discharge, particularly if wet or with invasive conductors.
- Fire risk: oxygen-enriched environment + ignition source (spark at pads/paddles) + fuel (drapes, alcohol prep). Minimise oxygen pooling; allow alcohol to dry.
- ECG leads and monitoring: ensure defib-proof leads; defib can saturate/damage equipment without proper protection.
Explain how a defibrillator generates the shock and relate selected energy to capacitor voltage.
Core circuit is a capacitor charged to a high voltage then rapidly discharged through the patient via switching circuitry.
- Energy stored in capacitor: E = 1/2 C V^2.
- Charge on capacitor: Q = C V; initial peak current approximately I0 = V0/R (R ≈ transthoracic impedance).
- If C is fixed, doubling selected energy requires increasing V by factor √2 (because E ∝ V^2).
- Actual delivered energy is less than stored due to internal resistance and waveform truncation; modern devices aim to deliver the selected energy to the patient.
What is transthoracic impedance (TTI) and why does it matter?
TTI is the effective resistance between electrodes across the chest; it strongly influences delivered current for a given selected energy.
- Ohm’s law: I = V/R. Higher TTI → lower peak current and reduced probability of defibrillation success.
- Typical adult TTI often ~50–100 Ω but can be higher with poor contact, obesity, hyperinflation etc.
- TTI is modifiable: pad size/position, gel, pressure, shaving hair, avoiding air gaps, end-expiration, repeated shocks.
Compare monophasic and biphasic defibrillation waveforms and explain why biphasic is more effective at lower energies.
Biphasic shocks reverse current direction part-way through, improving defibrillation efficiency and lowering defibrillation threshold.
- Monophasic: one direction of current; generally requires higher energies (e.g., 360 J) and causes more myocardial injury for similar success.
- Biphasic: two phases with polarity reversal; second phase helps neutralise residual membrane polarisation, reducing post-shock arrhythmogenicity and lowering DFT.
- Many biphasic devices use impedance compensation (adjust pulse width/voltage) to maintain effective current delivery across variable TTI.
A patient has very high chest impedance. What happens to the delivered current and what practical steps can you take?
High impedance reduces peak current and may reduce success; you should reduce impedance and ensure correct technique.
- Physics: I0 = V0/R, so as R increases, I decreases for a given voltage; myocardial current falls.
- Reduce TTI: ensure correct pad placement, use large pads, press firmly (paddles), add/replace gel, remove air gaps, shave hair, dry chest, remove patches.
- Consider AP pad position if appropriate; ensure pads not over bone prominence or device generator.
- If using biphasic, device may compensate; if persistent failure, escalate energy per algorithm/manufacturer guidance.
Why do larger defibrillation pads improve success and reduce burns?
They reduce transthoracic impedance and reduce current density at the skin.
- Larger area improves contact and lowers TTI → higher delivered current for the same selected energy.
- Current density is lower (conceptually J = I/A), reducing local heating and burn risk.
- Poor contact/air gaps concentrate current at edges, increasing burns; correct application is crucial.
Explain synchronised cardioversion and the R-on-T phenomenon in terms of cardiac electrophysiology and timing.
Synchronisation avoids delivering a shock during the vulnerable period of repolarisation.
- Shock during the T wave (relative refractory period) can induce VF (R-on-T).
- Synchronised mode detects R waves and triggers discharge shortly after the R wave to avoid the T wave.
- If sensing is poor (artefact, low amplitude), adjust lead selection/gain; if unstable and cannot synchronise, may need unsynchronised shock.
Previous FRCA-style calculation: A defibrillator uses a 100 microfarad capacitor. What voltage is required to store 200 J? What charge is stored?
Use E = 1/2 C V^2 and Q = C V.
- C = 100 microF = 100 × 10^-6 F = 1 × 10^-4 F.
- E = 1/2 C V^2 → V = √(2E/C) = √(400 / 1×10^-4) = √(4×10^6) = 2000 V.
- Q = C V = 1×10^-4 × 2000 = 0.2 C (200 mC).
Previous FRCA-style calculation: If the initial capacitor voltage is 2000 V and the patient impedance is 50 ohms, estimate the initial peak current. What if impedance is 100 ohms?
Use Ohm’s law for the initial condition (simplified).
- At 50 Ω: I0 = V0/R = 2000/50 = 40 A.
- At 100 Ω: I0 = 2000/100 = 20 A.
- This illustrates why high impedance reduces delivered current and can reduce defibrillation success at the same selected energy.
Previous FRCA-style viva: Describe factors that influence transthoracic impedance and how it changes during resuscitation.
TTI is affected by patient, electrode and technique factors and is not constant.
- Electrode factors: pad size (bigger lowers), gel quality, air gaps, position, paddle pressure, dried gel increases impedance.
- Patient factors: obesity, chest wall thickness, hyperinflated lungs, oedema, hair, sweating/wetness (complex effects—wetness can increase hazard and alter contact).
- Dynamic change: repeated shocks can reduce TTI (heating, improved contact); end-expiration may lower impedance.
Previous FRCA-style viva: Why can a shock cause skin burns even if the selected energy is not very high?
Burns relate to local current density and poor contact rather than just total energy.
- Small effective contact area or edge effects (air gaps) concentrate current → high current density → resistive heating.
- Dry gel/hair/poor pressure increases impedance and can create hot spots at the electrode-skin interface.
- Use large pads, good gel, firm pressure, avoid reusing dried pads, ensure full adhesion.
Previous FRCA-style viva: Explain the difference between defibrillation and cardioversion in terms of timing and energy delivery.
Both deliver electrical energy, but cardioversion is synchronised to avoid inducing VF.
- Defibrillation: unsynchronised shock for VF/pulseless VT; immediate delivery.
- Cardioversion: synchronised to R wave for organised tachyarrhythmias with a pulse (e.g., AF flutter SVT) to avoid R-on-T.
- Energy: cardioversion often starts lower and escalates; defibrillation uses higher energies per algorithm and device type (biphasic vs monophasic).
0 comments
Please log in to leave a comment.